Modified Levey-Jennings charts for calculated laboratory tests


Coskun A.

CLINICAL CHEMISTRY AND LABORATORY MEDICINE, vol.44, no.4, pp.387-390, 2006 (Journal Indexed in SCI) identifier identifier identifier

  • Publication Type: Article / Article
  • Volume: 44 Issue: 4
  • Publication Date: 2006
  • Doi Number: 10.1515/cclm.2006.066
  • Title of Journal : CLINICAL CHEMISTRY AND LABORATORY MEDICINE
  • Page Numbers: pp.387-390

Abstract

Background: In clinical laboratories, many test results such as low-density lipoproteins can be calculated using the results of specifically measured tests. However, quality control programs deal with only measured tests and there is no quality control procedure for calculated tests. I aimed to modify the Levey-Jennings chart for calculated tests as part of a total quality system.